DOC022.52.80205
2100AN
08/2012, Edition 2
User Manual
Seitenansicht 0
1 2 3 4 5 6 ... 59 60

Inhaltsverzeichnis

Seite 1 - User Manual

DOC022.52.802052100AN08/2012, Edition 2User Manual

Seite 2

In addition, two user-defined measurement units can be specified. Referto Application specific methods on page 45. The application specificmode of ope

Seite 3 - Table of Contents

Figure 3 Instrument components1 2100AN turbidimeter 7 StablCal® Calibration kit2 USEPA filter assembly 8 Gelex® secondary turbiditystandardiz

Seite 4

User interfaceFigure 4 Keypad1 ENTER key 7 FLOW key2 EDIT (arrow) keys 8 SIGNAL AVG key3 SAMPLE key 9 PRINT key4 RANGE key 10 LINE FE

Seite 5

Figure 5 Indicator lights1 Lamp icon light 7 PRINT light2 "CAL?" light 8 SIGNAL AVG light3 "Manual" light 9 CAL/Zero

Seite 6

StartupTurn the instrument on1. Put the instrument on a stable, level surface that is free of vibration.Do not put in direct sunlight.2. Make sure tha

Seite 7 - Specifications

calibrations. The manufacturer cannot guarantee the performance of the instrumentif calibrated with co-polymer styrenedivinylbenzene beads or other su

Seite 8 - General information

StablCal calibration procedure1. Remove the filterassembly. Refer to Change the filterassembly on page 48.2. Clean the lens of theUSEPA filter assem

Seite 9 - Product overview

StablCal standards storage• Do not move a StablCal standard to a different container for storage.Keep StablCal standards in the plastic case provided

Seite 10 - Product components

7. Push RATIO to turnRatio mode on.8. Put the stray lightstandard in the samplecell holder with thetriangle on the vialaligned with thereference mar

Seite 11 - Installation

7. Push RATIO toselect Ratio on or off.Ratio must be on forGelex standardsgreater than 40 NTU.For the 0–2 and 0–20 NTU Gelexstandards, select theRati

Seite 13 - English 11

Clean the sample cellC A U T I O N Chemical exposure hazard. Obey laboratory safety procedures andwear all of the personal protective equipment approp

Seite 14 - Standard operation

Indexing a single sample cellWhen measuring very low turbidity samples, use a single indexed sample cell or a flow cell for all measurements to get pr

Seite 15 - Calibration notes

7. Repeat step 6 untilthe lowest value isshown on the display.8. Put an orientationmark on the markingband near the top of thesample cell where thel

Seite 16 - 14 English

Matching sample cellsTo decrease the effects that optical differences among sample cells can have on turbidity, transmittance, color or absorbance mea

Seite 17 - Gelex notes

7. Repeat step 6 untilthe lowest value isshown on the display.8. Record the value.Put an orientation markon the marking bandnear the top of thesampl

Seite 18 - 16 English

Prepare dilution waterDilution water is used when indexing a sample cell or matching samplecells and to prepare formazin standards.1. Collect at least

Seite 19 - Prepare a sample cell

2. Put 1/2 to 2/3 of the sample cell into the ultrasonic bath and let it standuntil visible bubbles are removed.3. Remove the sample cell from the ult

Seite 20 - N O T I C E

Example: Measured value = 2450 NTUActual turbidity = 2450 × 5 = 12,250 NTUFigure 6 Prepare filtered sample using membrane or glass-fiberfilter1 Fil

Seite 21 - Indexing a single sample cell

Measurement• Measure samples immediately to prevent temperature changes andsettling. Before a measurement is taken, always make sure that thesample is

Seite 22 - 20 English

7. Read and record thevalue when stable.Note: To print or send(via RS232) ameasurement record,push PRINT.Absorbance and transmittance measurementMeas

Seite 23 - Matching sample cells

Table of ContentsSpecifications...

Seite 24 - 22 English

Absorbance and transmittance measurement procedureNote: To measure samples with negative absorbance, set the analytical zero using the sample with the

Seite 25 - Prepare the sample

7. Slowly put 250 mL ofthe sample in the inletreservoir.8. After the sampleflow stops and thedisplay stabilizes, readand record the value.Note: To p

Seite 26

Color measurement and calibration procedure1. Make sure that the455 nm filter assemblyis installed in theinstrument.2. Push RANGE toselect manual or

Seite 27 - W A R N I N G

7. Slowly put 250 mL ofdeionized water downthe interior edge of theinlet reservoir.Put the CU standarddown the interior edgeof the reservoir topreven

Seite 28

Measurement techniquesMeasurements may be made with different operation mode settings andoptional accessories.Calibrate the instrument whenever the sa

Seite 29 - Measurement notes

if interferences caused by color or light absorbing particles are notpresent.Using the air purge systemThe air purge system is used to keep condensati

Seite 30 - 28 English

2. Fill the flow cell and tubing with water and make sure that there areno leaks or air bubbles.Note: Air bubbles collect in areas that are not cleane

Seite 31 - Color measurement

• For short-term storage (a few hours), flush the system with distilled ordeionized water and leave the flow cell full of the flush water tominimize a

Seite 32 - 30 English

Use the flow cell specifications in Table 3 to calculate the correct fill time.Make sure that the fill time includes time to fill the system and to fu

Seite 33 - English 31

7. Push ENTER.The display shows"MM-SS MEA" (or anactual measurementtime if a measurementtime has been selectedpreviously).8. Push the arro

Seite 34 - Measurement techniques

Clean the sample cell...

Seite 35

Install a cell adapterNote: Use the application specific calibration (ASC) ability of the instrument toprovide direct reading of results with cell ada

Seite 36

Connect to a printer or computerUse the serial interface (RS232) connector on the back of the instrumentto transmit data from the instrument to an ext

Seite 37

Table 5 RS232 command set (continued)Command DescriptionRMN Gets the recorder minimum value.To change the recorder minimum value, enter RMN=XXXXX(XXX

Seite 38

Prepare formazin standardsFor the best accuracy and long-term data comparability, use formazinstock solution from Hach to make formazin standards.Note

Seite 39 - Use a cell adapter

• Store the oiling cloth in a plastic storage bag to keep the cloth clean.• If power is lost during calibration, the new calibration data is lost andt

Seite 40

7. Apply a small beadof silicone oil from thetop to the bottom of thesample cell.8. Use the oiling clothprovided to apply the oilequally to the surf

Seite 41 - Computer (RS232) commands

Making 4000-NTU formazin stock solutionW A R N I N G Chemical exposure hazard. Obey laboratory safety procedures and wear all ofthe personal protectiv

Seite 42 - Advanced operation

Special research applicationsThe instrument has special features and operations for special researchapplications.Application specific methodsUse the a

Seite 43 - Prepare formazin standards

Either ASC can be changed at any time, so recalibration is notnecessary.The sample is under-range if the display flashes 0s.If the display flashes 0s

Seite 44

4. Push UNITS/Exit to return to measurement mode.Change an ASC unit name or one ASC data pointAny ASC unit name or data point can be changed.Note: Pus

Seite 45 - English 43

Using a manual flow cell...

Seite 46

Delete all the ASC data pointsEither of the ASC calibration curves can be deleted and the ASC unitname changed back to the factory default.1. Push UNI

Seite 47 - Application specific methods

• Replace the lamp with the same size, style and electrical rating(4708900). Refer to Replacement parts and accessorieson page 53.• Do not touch the l

Seite 49 - Delete one ASC data point

Replace a fuseD A N G E R Fire hazard. Use the same type and current rating to replace fuses.Replacement parts:• Fuse for 115 V operation, time-delay,

Seite 50 - Maintenance

Table 8 Error codes (continued)Error Description SolutionERR03 Low light error1. Put the sample in the instrument again.2. Make sure that the lamp ic

Seite 51

Table 9 Diagnostic codes (continued)Code Display Description23 Test results are shown. Keyboard test24 Test results are shown. Memory testDelete ca

Seite 52 - 50 English

Replacement parts and accessories (continued)Description Quantity Item no.Power cord, European, 230 VAC, VDEapproved1 4683600 Printer paper, thermal

Seite 53 - Troubleshooting

Replacement parts and accessories (continued)Description Quantity Item no.Pump, vacuum/pressure, 115V, 60 Hz,1.2 cfm1 2424800 Pump, vacuum/pressure,

Seite 56 - 54 English

Delete all the ASC data points...

Seite 57 - English 55

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Seite 58 - 56 English

SpecificationsSpecifications are subject to change without notice.Specification DetailsMeasurement method NephelometricRegulatory Meets EPA Method 180

Seite 59

Specification DetailsOperating conditions Temperature: 0 to 40 °C (32 to 104 °F)Relative humidity: 0–90% at 25 °C, 0–75% at 40 °C,noncondensingAltitud

Seite 60

C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderateinjury.N O T I C E Indicates a situation which, if not a