Section 8 Troubleshooting
8.1 General problems and GSM module errors
Table 40 General Problems
Problem Cause/Solution
Incorrect display reaction when the
display is touched
Calibrate touch screen with finger or stylus. If this is not possible: Start factory
calibration through web access.
No trend line available. Configure the logger in the appropriate probe.
Communication problems
Check the probe plug, check probe cable damage, check network plug and cable at
sc1000 networks.
LED indicator in the probe module
blinks red
Refer to Communication problems
LED indicator in the probe module is
turned off
Check the fuses, if the probe module is connected to power supply.
Missing measured values after
changing probe (mA output card,
fieldbus card)
New configuration for the output cards is necessary. Configure the new probe with its
serial number. After that, delete the not used probe in the device management.
Local web access is not available
Check the Ethernet connection, LAN configuration and IP address in the
sc1000 SETUP, BROWSER ACCESS menu.
Local web access is blocked with the
“MENU ACCESS” message
The display module is not in the “Measured value display” mode.
Local web access is blocked,
missing password
Activate password in the sc1000 SETUP, SECURITY SETUP menu (refer to
section 6.3.10, page 115).
Display module operation denied
with the “WEB ACCESS” message
Close the external web access to enable local display operation again.
General probe specific error
messages
Check the probes in the SENSOR DIAGNOSTIC menu on error messages and
warnings. Errors are indicated with a red background in the according displayed
measurement value.
Screen is blue, no measurements
are displayed
Check if the probes are connected. If the probes are connected, scan for new
devices. Check the measurement screen configuration. If no device is configured add
it to the measurement screen configuration.
New devices (probes, modules) are
connected to the sc1000 controller
but were not listed during initial bus
scan
Check whether missing devices are connected to local or – when installed – to
remote probe modules. Use the serial numbers for identification.
For remote probe modules check proper termination of all network cables.
For local probe modules retry with interchanged device connectors.
Refer to Communication problems.